BIOMECHANICAL EVALUATION OF HYDROXYAPATITE -POLYETHYLENE COMPOSITE COATING
CSTR:
Author:
Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    To evaluate the biomechanical differences of Hydroxyapatite-Polyethylene composite coating and uncoated Ti-alloy implants. Methods The canine transcortical implant model was utilized At the end of the implantation periods push-out tests were carried out to acquire bone-coating interface shear strength of hydroxyapatite-polyethylene coating with uncoated Ti-alloy implants as control. Results The hydroxyapatite-polyethylene coating in every group showed greater shear strengths than uncoated implants being 3-5 times as much. The difference was significant statistically. (P<0.01) The strength increased with time reached the peak value at 12 weeks. Conclusions The interface shear strength of hydroxyapatite-polyethylene composite coating was much greater than that of the uncoated implants the results have revealed its good biomechanical characteristics which provided an effective method of biological fixation.

    Reference
    Related
    Cited by
Get Citation

. BIOMECHANICAL EVALUATION OF HYDROXYAPATITE -POLYETHYLENE COMPOSITE COATING[J]. Journal of medical biomechanics,2001,16(3):179-172

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:
  • Revised:
  • Adopted:
  • Online:
  • Published:
Article QR Code